Product Info

Table Of Contents
High sensor settings
Item Default
setting
Limits Increments
Alert on high Off - -
Time before high 15 minutes 5 to 30 minutes 5 minutes
Rise Alert Off - -
Rise Limit Two up
arrows
1 up arrow
(1 mg/dL/min)
2 up arrows
(2 mg/dL/min)
3 up arrows
(3 mg/dL/min)
•Custom limit
(1.0 to 5.0 mg/dL/min)
High Snooze 1 hour 5 minutes to 3 hours 5 minutes
Low sensor settings
Item Default
setting
Limits Increments
Low SG alert
limit
60 mg/dL 50 to 90 mg/dL 5 mg/dL
Suspend before
low
Off - -
Suspend on low Off - -
Alert before low Off - -
Alert on low Off - -
Low Snooze 20 minutes 5 minutes to 1 hour 5 minutes
Resume basal
alert
Off - -
Product specifications and safety information
Product specifications and safety information 313