Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Percent temp basal
The default value is 100 percent of basal programming. For example, if you
program six units of basal per day, the default temp basal will be six units per day.
Range 0 to 200%
Default 100% of basal programming
Increment 5%
Program safety checks
A single fault condition will cause the pump to suspend insulin delivery. Maximum
infusion with a single fault condition is 0.2 units.
Pump dimensions
The pump dimensions in inches are no greater than 3.78 length x 2.11 width x
0.96 depth.
The pump dimensions in centimeters are no greater than 9.60 length x 5.36 width
x 2.44 depth.
Pump memory
User settings and pump history are stored in non-volatile memory which will retain
data. The memory size will hold 90 days of pump history before it becomes full
and has to be written over. The viewable history on the pump is 30 days. This
information can be accessed on the History screen.
Pump weight
The mass of the insulin pump without battery and consumables is less than 106
grams.
Sensor default settings
High sensor settings
Item Default
setting
Limits Increments
High SG alert
limit
250 mg/dL 100 to 400 mg/dL 5 mg/dL
Alert before high Off - -
312 Chapter 16