Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Alert range Increment Default
value
First reminder occurs at 5 to 50 units. Second reminder
occurs at 50 percent of the remaining specified amount.
The second reminder is automatic and cannot be changed
by the user.
1 unit 20 units
Max Bolus
Range 0 to 25 units
Default 10 units
Normal bolus
Range is 0.025 to 25 units of insulin, and limited by the Max Bolus setting.
Occlusion detection
When occlusion is detected, the Insulin flow blocked alarm occurs. The occlusion
alarm is triggered by an average of 2.23 units of missed insulin (standard bolus) or
1.97 units of missed insulin (quick bolus). The MiniMed 770G insulin pump is
intended for use with U-100 insulin. This table shows occlusion detection for four
different situations when using U-100 insulin.
Rate Minimum time
before alarm
Average time
before alarm
Maximum time
before alarm
bolus delivery (10 units at
standard speed)
71 seconds 95 seconds 136 seconds
bolus delivery (10 units at
quick speed)
9 seconds 10 seconds 14 seconds
basal delivery (1.0 u/h) 2.00 hours 2.50 hours 3.80 hours
basal delivery (0.025 u/h) 123.38 hours 142.03 hours 178.33 hours
Note: Certain factors, such as ambient temperature changes or the
presence of air in the infusion set or the reservoir, can delay an occlusion
alarm.
■