Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Vibrate mode range 0 to 20 increments or Max Bolus limit, whichever
comes first
Default step size 0.1 unit
Adjustable step size 0.1 to 2 units per increment up to Max Bolus limit
Environmental conditions
The MiniMed 770G insulin pump system is designed to withstand most conditions
encountered in your daily life. For more details about environmental conditions,
such as exposure to magnetic fields and radiation, waterproof capabilities, and
extreme temperatures, see User safety, on page 7.
• Pump storage temperature range without a AA battery is from -4°F (-20°C) to
122°F (50°C).
• Pump operating temperature range is from 41°F (5°C) to 104°F (40°C).
• Operating air pressure range is from 10.2 psi (700 hPa) to 15.4 psi (1060 hPa).
• Storage air pressure range is from 7.2 psi (496.4 hPa) to 15.4 psi (1060 hPa)
• Relative humidity (RH) range during operation is from 20% to 90%.
• RH range during storage is from 5% to 95%.
Essential performance
In the specified environmental conditions, the insulin pump maintains essential
performance of the following functionalities:
• Delivery accuracy
• Occlusion detection
• Empty reservoir detection
• Detection of power loss
• Pump therapy status
• Notification annunciation and display
Filling the infusion set and cannula
• The cannula can be filled from 0.025 units to 5.1 units, in increments of
0.025 units.
• The standard fill rate is 1.5 units per minute.