Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
• The maximum infusion pressure generated and the occlusion threshold
pressure using a 3.0-mL reservoir is 13.15 psi (90.67 kPa). The average resulting
bolus volume generated upon clearing the occlusion is 0.0112 mL (equivalent
to 1.12 units of U-100 insulin).
• The following image is a representative delivery accuracy curve. The Trumpet
Curve represents the maximum percentage change from the expected insulin
dosage for a given time interval, known as the observation window, during
the infusion of insulin. The upper curve corresponds to positive changes, and
the lower curve corresponds to negative changes.
0 10 20 30 40 50 60 70 80 90 100
−20
−15
−10
−5
0
5
10
15
20
25
Trumpet Curve at intermediate rate of 1 U/h
Observation Window [Minutes]
Percent Error [%]
Overall Error
Maximum Error
Minimum Error
Easy Bolus feature
The Easy Bolus feature lets the user set up and deliver a Normal Bolus when the
pump is in Sleep Mode. This is done using
and with the help of audio and
vibration cues.
Audio mode range 0 to 20 increments or Max Bolus limit, whichever
comes first
■
Product specifications and safety information
Product specifications and safety information 307