Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
2. Place the battery cap onto the pump. Use the bottom edge of the pump clip
to turn the cap to the right and tighten.
CAUTION: Do not overtighten or undertighten the battery cap.
Overtightening the battery cap can cause damage to your pump
case. Undertightening the battery cap will prevent the pump from
recognizing the new battery. Turn the battery cap clockwise until
the cap is aligned horizontally with the pump case, as shown in
the following example.
Note: If this is the first time you have inserted a battery in your pump,
the Startup Wizard begins. For more information about the Startup
Wizard, see Entering your startup settings, on page 41. If this is not the first
time you have inserted a battery into your pump, the Home screen
appears and the pump resumes your basal delivery.
Removing the battery
CAUTION: Do not remove the battery unless you insert a new battery
or store the pump. Your pump cannot deliver insulin while the battery
is removed. After you remove an old battery, be sure to replace it with
a new battery within 10 minutes to clear the Insert battery alarm and
avoid a Power loss alarm. If power loss occurs, you must re-enter your
time and date settings.
■
First steps
First steps 39