Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Bolus delivery
Bolus Speed options • Standard: 1.5 units/minute
• Quick: 15 units/minute
Bolus programming increments • 0.025 units
•0.05 units
•0.1 units
Fluid delivered/stroke • 0.25 μL (microliter) for 0.025 unit pump stroke
• 0.5 μL for 0.05 unit pump stroke
• 2.0 μL for 0.2 unit pump stroke
Bolus Wizard feature default settings
Item Default Limits Increments
Carb units grams - -
Insulin to carb ratio None 1–200 g/u 0.1 g/u for 1–9.9 g/u;
1 g/u for ratios of 10 g/u to
200 g/u
Insulin Sensitivity Factor None 5–400 mg/dL 1 mg/dL
BG Target None 60–250 mg/dL 1 mg/dL
Active Insulin Time 4 hours 2 to 8 hours 15 minutes
Bolus Wizard feature specifications
There are four different formulas the Bolus Wizard feature uses to estimate a bolus,
depending on your current BG. The following formulas apply only when the carb
units are in grams.
1. If your current BG is greater than your High BG Target, the Bolus Wizard
feature subtracts active insulin from the BG correction estimate, then adds this
to the food estimate to get the total bolus estimate. However, if the result of
subtracting active insulin from BG correction estimate is a negative number
(less than zero), the total bolus estimate is based only on the food estimate.
■
Product specifications and safety information
Product specifications and safety information 303