buttons. Remove the pump and begin using your backup insulin plan per the
direction of your healthcare professional. If the pump is accidentally
programmed while the screen is broken or unreadable, this could result in
high or low BG levels. If your screen is damaged, contact 24-Hour Technical
Support to arrange for shipment of a replacement pump.
Only use rapid-acting U-100 insulin (Humalog and NovoLog) that has been
prescribed by your healthcare professional for use with an infusion pump. Do
not put any other drugs or medications inside your reservoir for use with this
pump. Other drugs or medications are not intended for use with this pump.
Use of other drugs or medications can cause serious injury.
Always make sure the infusion set is disconnected from your body before you
rewind your pump or fill the infusion set tubing. Never insert the reservoir into
the pump while the tubing is connected to your body. Doing so could result
in an accidental infusion of insulin.
Do not insert the reservoir in the pump if you did not rewind your pump.
Doing so could result in an accidental infusion of insulin.
Do not use the MiniMed 770G insulin pump or additional system devices
adjacent to other electrical equipment which may cause interference with the
normal system operation. This includes mobile communication devices such
as cell phones that are not paired with the MiniMed 770G System, GPS
navigation systems, anti-theft systems, and any electrical equipment that has
an output transmitter power greater than 1W. For more information about
recommended separation distance guidelines between the insulin pump and
common RF emitters, see Guidance and manufacturer's declaration, on
page 317. The recommended separation distance between the insulin pump
and common RF emitters is 12 inches. Other electrical equipment that may
compromise normal system operation has been contraindicated. For more
information, see Exposure to magnetic fields and radiation, on page 24.
Do not unscrew or retighten the tubing connector on the reservoir while the
infusion set is connected to your body. Doing so could result in an accidental
infusion of insulin.