Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Basal delivery
Delivery rate range 0 to 35 units per hour or the Max Basal Rate
amount, whichever is lower.
Max Basal Rate default 2 units per hour
Basal patterns Maximum of 8 patterns. Each pattern covers a 24-
hour period and can have up to 48 rates. Rates are
set in 30-minute increments.
Basal pattern names Fixed names: Basal 1, Basal 2, Basal 3, Basal 4, Basal 5,
Workday, Day Off, Sick Day
Increments • 0.025 units per hour for basal amounts in the
range 0 to 0.975 units
• 0.05 units per hour for basal amounts in the range
1 to 9.95 units
• 0.1 units per hour for basal amounts of 10 to
35 units
BG Target
Maximum targets 8
Range 60 to 250 mg/dL
Default value for High blood
glucose (BG) targets and Low
BG targets
None
Note: Auto Mode uses a fixed BG Target of 150 mg/dL.
BG meter value
The most recent BG value received from the meter. If you are using an Accu-Chek
Guide Link meter, this value appears on the Home screen when the Sensor feature
is off. This value also appears in the Bolus Wizard screen when setting up a bolus.
Expiration 12 minutes
Range 20 to 600 mg/dL
302 Chapter 16