How-To Guide

Option Silences these alerts
High & Low
Alerts
Alert on high, Alert before high, Rise Alert, Alert on low, Alert before
low, Suspend before low, and Resume Basal Alert
Note: Alert on low cannot be silenced if the
SmartGuard Suspend on low or SmartGuard Suspend
before low options are turned on.
All Sensor
Alerts
All of the alerts listed previously for High & Low Alerts, plus the
following:
All calibration alerts, reminders, or error messages
All alerts relating to sensor insertion, including alerts about sensor
warm-up, changing your sensor, sensor expiration, sensor errors,
connection issues, and so on
All alerts related to your transmitter, including all alerts about your
transmitter battery and all connection issues
To silence Glucose alerts:
1. Press
and go to the Alert Silence screen.
Audio Options > Alert Silence Options
2. Select High Alerts Only, High & Low Alerts, or All Sensor Alerts to set the
alerts you want silenced. Refer to the previous table for details about the
alerts silenced with each selection.
Note: If you select All Sensor Alerts, you will not receive most alerts
related to your SG readings, your sensor, calibration requirements, or
your transmitter. The Low SG XX mg/dL (XX represents 50 mg/dL or