Title and text Type Explanation Next steps
Suspend on low
Delivery stopped.
Sensor glucose
XXX mg/dL.
Check BG.
Alarm Your SG value is
at or below the
low limit you
specified.
Suspend on low
is not available in
Auto Mode.
• Select OK to clear the alarm.
• Check your BG. If necessary, treat
your BG as directed by your
healthcare professional.
Transmitter
battery depleted
Recharge
transmitter now.
Alert The battery in
the transmitter
needs to be
recharged. SG
values are not
recorded or
transmitted until
you recharge
transmitter.
• Select OK to clear the alert.
• Recharge your transmitter.
SmartGuard Auto Mode alerts and messages
The following table lists the most common or serious alerts and messages related
to Auto Mode. The table also explains the meaning, consequences, and the
reasons why these notifications appear, and provides any necessary steps for
problem resolution.
Title and text Type Explanation Next steps
Auto Mode
started
Current action
canceled.
Alert This alert
happens when
the user starts an
operation that is
not allowed in
Auto Mode while
the pump is
transitioning to
Auto Mode.
• Select OK to clear the alert.
• Allow your pump to complete its
transition to Auto Mode.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 273