How-To Guide

Title and text Type Explanation Next steps
Sensor
connected
If new sensor,
select Start New.
If not, select
Reconnect.
Message The transmitter
has detected
that you have
connected a
sensor. The
pump needs to
know if this is a
new sensor or if
you have
reconnected
your old sensor.
If you have connected a new
sensor, select Start New Sensor.
If you have reconnected a sensor
you have been using, select
Reconnect Sensor.
In either case, a "Warm-up"
message appears on your Home
screen, and you will be
prompted to enter a BG value
when your sensor is ready for
calibration. Your pump starts
receiving your SG values again
after the two-hour initialization is
complete.
Sensor
connected
Start new sensor.
Message The pump has
detected that
this is a new
sensor, which
needs to be
started and
warmed-up.
Select Start New Sensor.
The alert will close and a "Warm-
up" message appears on the
sensor graph with a progress bar.
Sensor expired
Insert new
sensor.
Alert The sensor has
reached the end
of its useful life.
Change your sensor. For details,
see your sensor user guide.
Select OK to clear the alert.
Sensor signal
not found
See User Guide.
Alert After multiple
attempts, the
pump failed to
detect the
transmitter and is
unable to receive
sensor signal.
Select OK to clear the alert.
If your pump still cannot find the
sensor signal, call 24-Hour
Technical Support for assistance.
Alarms, alerts, and messages