Title and text Type Explanation Next steps
Change sensor
Sensor not
working properly.
Insert new
sensor.
Alert This alert occurs
when the
transmitter
diagnoses a
problem with the
sensor that
cannot be
resolved.
• Select OK to clear the alert.
• Change your sensor. For details,
see your sensor user guide.
Check
connection
Ensure
transmitter and
sensor
connection is
secure, then
select OK.
Alert The pump fails
to detect the
transmitter and is
unable to receive
sensor signal.
• Select OK to clear the alert.
• If your sensor is fully inserted,
select Yes. If your sensor is not
fully inserted, select No.
• If your sensor was not fully
inserted, insert a new sensor.
• If you still cannot connect your
sensor, see My pump cannot find
the sensor signal, on page 288.
Lost sensor
signal
Move Pump
closer to
transmitter. May
take 15 minutes
to find signal.
Alert Transmitter signal
has not been
received for
30 minutes
during or after
initialization.
• Move your pump closer to your
transmitter. It can take up to
15 minutes for your pump to
start communicating with your
transmitter.
• Select OK to clear the alert.
Low battery
transmitter
Recharge
transmitter
within 24 hours.
Alert The battery in
the transmitter
needs to be
recharged within
24 hours.
• Select OK to clear the alert.
• Recharge your transmitter as
soon as possible.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 267