How-To Guide

Title and text Type Explanation Next steps
Resume Dual
bolus?
XX of YY U
delivered.
Resume
delivery of ZZ U
now, and
AA U Square for
XX:XX hr?
Message The Now portion of a Dual
Wave bolus delivery has
been interrupted because
the pump battery was
removed. If it is within
10 minutes since this
interruption, you can
resume this bolus.
Check the message to
see how much of the
Dual Wave bolus was
actually delivered.
To cancel remaining
amount of bolus, select
Cancel.
To resume remaining
amount of bolus, select
Resume.
Resume Square
bolus?
XX of YY U
delivered for
XX:XX hr.
Resume
delivery of ZZ U
for XX:XX hr?
Message The Square Wave bolus
delivery was interrupted. If
it is within 10 minutes
since this interruption, you
can resume this bolus.
Check the message to
see how much of the
Square Wave bolus was
actually delivered.
To cancel remaining
amount of bolus, select
Cancel.
To resume remaining
amount of bolus, select
Resume.
Rewind
required
Delivery
stopped.
Rewind was
required due to
pump error.
Select OK to
continue. See
User Guide.
Alarm Your pump encountered
an error.
Select OK to clear the
alarm after the pump
has completed
rewinding.
Select Reservoir &
Tubing from the Home
screen to start the new
reservoir process using a
new infusion set and
reservoir. For details, see
Setting up the reservoir
and infusion set, on
page 117.
Alarms, alerts, and messages
Alarms, alerts, and messages 261