Title and text Type Explanation Next steps
Reservoir
estimate at 0 U
To ensure
insulin delivery,
change
reservoir.
Alert Your reservoir level is
estimated at 0 units.
• Select OK to clear the
alert.
• Change the reservoir
now.
Resume bolus?
XXX of YYY U
delivered.
Resume
delivery of
ZZZ U?
Message A normal bolus delivery has
been interrupted because
the pump battery was
removed. If it is within
10 minutes since this
interruption, you can
resume this bolus.
• Check the message to
see how much of the
bolus was actually
delivered.
• To cancel remaining
amount of bolus, select
Cancel.
• To resume remaining
amount of bolus, select
Resume.
Resume Dual
bolus?
XX of YY U
delivered.
Resume
delivery of ZZ U
for XX:XX hr?
Message The Square portion of Dual
Bolus delivery has been
interrupted. If it is within
10 minutes since this
interruption, you can
resume this bolus.
• Check the message to
see how much of the
Dual Wave bolus was
actually delivered.
• To cancel remaining
amount of bolus, select
Cancel.
• To resume remaining
amount of bolus, select
Resume.
260 Chapter 13