Title and text Type Explanation Next steps
Pump restarted
Delivery
stopped.
Settings
unchanged.
Select OK to
continue. See
User Guide.
Alarm Your pump has
encountered a problem
and has restarted. Your
settings have not been
changed.
• Select OK to continue.
•If the pump was
delivering a bolus or
filling the cannula, check
Daily History and re-
evaluate your need for
insulin.
• If this alarm recurs
frequently, write down
the error code displayed
on the alarm screen (you
can also find it in your
Alarm History) and call
24-Hour Technical
Support.
Replace battery
Battery life less
than 30
minutes. To
ensure insulin
delivery, replace
battery now.
Alert Battery life is low and will
be exhausted within 30
minutes.
• Select OK to clear the
alert.
• Replace the AA battery.
Replace battery
now
Delivery
stopped.
Battery must be
replaced to
resume
delivery.
Alarm Insulin delivery has stopped
due to low power. Battery
was not replaced after the
Low battery Pump alert.
Replace the battery
immediately to resume
insulin delivery.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 259