How-To Guide

Title and text Type Explanation Next steps
Pump error
Delivery
stopped.
Settings
unchanged.
Pump restart
needed. Select
OK to restart.
See User Guide.
Alarm A pump error has occurred,
you need to restart your
pump.
Select OK to restart your
pump.
•If the pump was
delivering a bolus or
filling the cannula, check
Daily History and
re-evaluate your need
for insulin.
If this alarm recurs
frequently, write down
the error code displayed
on the alarm screen (you
can also find it in your
Alarm History) and call
24-Hour Technical
Support.
Pump error
Delivery
stopped.
Settings
unchanged.
Select OK to
continue. See
User Guide.
Alarm Your pump encountered
an error but a restart is not
necessary. The issue is
resolved. Your settings are
not changed.
Select OK to resume
basal delivery.
•If the pump was
delivering a bolus or
filling the cannula, check
Daily History and
re-evaluate your need
for insulin.
If this alarm recurs
frequently, write down
the error code displayed
on the alarm screen (you
can also find it in your
Alarm History) and call
24-Hour Technical
Support.
258 Chapter 13