Title and text Type Explanation Next steps
Pump error
Delivery
stopped.
Current settings
cleared. Pump
restart needed.
Select OK to
restart and then
re-enter your
settings. See
User Guide.
Alarm Your pump encountered
an error and will restart.
Your pump settings will
return to factory default
values.
• Select OK to restart your
pump.
• When the pump restarts,
follow instructions on
the pump display.
• After restart, check
settings and re-enter
values as needed.
• If you recently saved
backup settings in
Manage Settings, use
Restore Settings.
•If the pump was
delivering a bolus or
filling the cannula, check
Daily History and re-
evaluate your need for
insulin.
• If this alarm recurs
frequently, write down
the error code displayed
on the alarm screen (you
can also find it in your
Alarm History) and call
24-Hour Technical
Support.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 257