How-To Guide

Title and text Type Explanation Next steps
No reservoir
detected
Rewind before
loading
reservoir.
Alarm There is no reservoir in the
pump or the reservoir is
not properly locked into
place.
Select Rewind.
Ensure that your
reservoir is filled with
insulin.
When prompted, ensure
that your reservoir is
inserted and properly
locked into place.
Power error
detected
Delivery
stopped.
Record your
settings by
uploading to
CareLink or
write your
settings on
paper. See User
Guide.
Alarm The internal power source
in your pump is unable to
charge. Your pump is
operating on the AA
battery only.
Select OK to clear the
alarm.
Check your BG and treat
as necessary.
Record your settings as
soon as possible
because your AA battery
may not last long.
Call 24-Hour Technical
Support for assistance
with your pump.
Power loss
AA battery was
removed for
more than 10
min or power
was lost. Select
OK to re-enter
time and date.
Alarm Your pump battery has
been out for more than
ten minutes, and your
pump has lost power. You
must reset your time and
date.
Select OK to go to the
Time & Date screen.
Enter the current time,
time format, and date.
256 Chapter 13