Title and text Type Explanation Next steps
Manage
settings error
Delivery
stopped.
Backup settings
cleared from
Manage
Settings.
Current settings
are working
properly. Select
OK to restart.
See User Guide.
Alarm A pump error has occurred,
and you need to restart
your pump. Your backup
settings have been lost, but
your current settings are
unchanged.
• Select OK to restart your
pump. Your current
settings are unchanged.
Only your backup
settings are lost.
• When the pump restarts,
follow instructions on
the pump display.
•If the pump was
delivering a bolus or
filling the cannula, check
Daily History and
evaluate your need for
insulin.
Max Fill
reached
3X.X U. Did you
see drops at
the end of
tubing?
Alarm You have exceeded the
number of units expected
to fill the tubing. By now,
insulin should be at the
end of the tubing.
• If you see drops at the
end of the tubing, select
Yes.
• If you do not see drops,
select No.
• Follow instructions
displayed on the pump.
Max Fill
reached
4X.X U. Remove
reservoir and
select Rewind
to restart New
Reservoir
procedure.
Alarm You have exceeded the
number of units expected
to fill the tubing. By now,
insulin should be at the
end of the tubing.
• Remove the reservoir.
• Check if you still have
insulin in the reservoir. If
you do, you can
continue using the same
reservoir.
• Select Rewind to restart
the new reservoir
procedure.
■
Alarms, alerts, and messages