Title and text Type Explanation Next steps
Insulin flow
blocked
Check BG.
Consider
testing ketones.
Estimated 0 U
insulin in
reservoir.
Change
reservoir and
infusion set.
Alarm Your pump has detected
that the insulin flow was
blocked and there is no
insulin in the reservoir.
• Check your BG. Consider
checking ketones and
take an injection if
needed.
• Remove your infusion
set and reservoir.
• Select Rewind to start
the new reservoir
process using a new
infusion set and
reservoir.
If a bolus delivery was in
progress when the alarm
occurred:
•Check the Daily History
screen for the amount of
bolus already delivered
before the pump
alarmed.
• Consider delivering
remaining bolus, if the
bolus insulin was not
included in an insulin
injection.
252 Chapter 13