Title and text Type Explanation Next steps
Insulin flow
blocked
Check BG.
Consider
testing ketones.
Check reservoir
and infusion
set.
Alarm Your pump has detected
that the basal or bolus
insulin flow was blocked.
• Check your BG. Consider
checking ketones and
take an injection if
needed.
• Remove your infusion
set and reservoir.
• Select Rewind to start
the new reservoir
process using a new
infusion set and
reservoir.
If a bolus delivery was in
progress when the alarm
occurred:
•Check the Daily History
screen for the amount of
bolus already delivered
before the pump
alarmed.
• Consider delivering
remaining bolus, if the
bolus insulin was not
included in an insulin
injection.
WARNING: Do not use Auto Mode for a period of time after
giving a manual injection of insulin by syringe or pen. Manual
injections are not accounted for in Auto Mode. Therefore, Auto
Mode could deliver too much insulin. Too much insulin may
cause hypoglycemia. Consult with your healthcare professional
for how long you need to wait after a manual injection of insulin
before you resume Auto Mode.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 251