How-To Guide

Title and text Type Explanation Next steps
Check settings
Startup Wizard
settings
complete.
Check and set
up your other
settings.
Alert Some settings have been
cleared or reverted to
factory default values.
Select OK to clear the
alert.
Review any settings that
you have not already set
in Startup Wizard and re-
enter the values, if
necessary.
Critical pump
error
Delivery
stopped. Pump
not working
properly. Stop
using pump.
Remove
infusion set
from body.
Consider other
insulin
treatment. See
User Guide.
Alarm Your pump has
encountered an error that
cannot be resolved. For
example, your pump may
have a mechanical
problem.
The pump is not able to
deliver insulin. Remove
your infusion set and stop
using your pump.
Consider another form
of insulin delivery.
Check your BG, and treat
as necessary.
Write down the error
code that appears on
the alarm screen.
Call 24-Hour Technical
Support for assistance
with your pump.
Delivery limit
exceeded
Delivery
stopped. Check
BG. See User
Guide for more
information.
Alarm Your pump has suspended
because the hourly delivery
limit was met. This limit is
based on the maximum
bolus and maximum basal
setting. If this alarm occurs
during a bolus, the bolus is
canceled before it can
complete.
Check your BG.
Select Resume Basal.
Check Bolus History and
re-evaluate your need
for insulin.
Continue to monitor
your BG.
Alarms, alerts, and messages
Alarms, alerts, and messages 247