Title and text Type Explanation Next steps
Battery failed
Insert a new AA
battery.
Alarm The pump battery does not
have enough power.
• Select OK to clear the
alarm.
• Remove the old battery
and insert a new AA
battery.
Battery not
compatible.
See User Guide.
Alarm The battery that you
inserted into the pump is
not compatible.
• To clear the alarm,
remove the
incompatible battery.
• Insert a new AA battery.
Bolus not
delivered
Bolus entry
timed out
before delivery.
If bolus
intended, enter
values again.
Alert Bolus values entered, but
bolus was not delivered
within 30 seconds.
• Select OK to clear the
alert.
•If bolus delivery was
intended, check your BG,
re-enter bolus values
and deliver bolus.
Bolus stopped
Cannot resume
bolus or
cannula fill.
XX.XXX of
YY.YYY U
delivered.
ZZ.ZZZ U not
delivered. If
needed, enter
values again.
Alarm The battery power was
exhausted while a bolus or
Fill Cannula was in
progress, or you did not
respond to the Resume
bolus? message after
replacing the battery.
• Note the amount of
insulin not delivered.
• Replace the AA battery.
• Select OK to clear the
alarm.
• Deliver the remaining
bolus amount if needed.
246 Chapter 13