How-To Guide

Line Item Instructions
Auto Mode turned off
Turn on Auto Mode in the SmartGuard, Auto
Mode screen.
Sensor not ready
Do the following:
Check to see if your pump has a transmitter
ID entered in Utilities, Device Options. For
example, GT6133333M.
Make sure your pump is paired with a
transmitter. For more information, see
Pairing your pump and transmitter, on
page 199.
Check your Home screen. If you see
,
move your pump and transmitter closer
together. The pump will try to find the
transmitter signal.
If after 30 minutes the pump and
transmitter are still not communicating, you
will receive a Lost sensor signal alert. Check
that the sensor is still inserted in the skin,
and the transmitter and sensor are still
connected. Move your pump closer to your
transmitter.
If your SG is outside of the 40 to 400 mg/dL
range, your pump will not enter Auto Mode.
Sensor off
Turn on the sensor in the Utilities, Sensor
Settings screen.
Bolus in progress
Wait until the bolus is complete or stop the
bolus yourself before Auto Mode can activate.
Delivery suspended
If insulin delivery is suspended, Auto Mode
cannot activate. Treat low BG if necessary as
instructed by your healthcare professional.
224 Chapter 12