Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The GST3C transmitter is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the GST3C transmitter should assure that it is
used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Envi-
ronment - Guidance
Voltage dips, short in-
terruptions and volt-
age variations on
power supply lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
Not applicable Requirement does not ap-
ply to this battery powered
device.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 400 A/m
4000 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical lo-
cation in a typical domestic,
commercial, or hospital
environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The GST3C transmitter is intended for use in the electromagnetic environment speci-
fied below. The customer or user of the GST3C transmitter should assure that it is
used in such an electromagnetic environment.
Immunity Test IEC 60601 Level Compliance Lev-
el
Electromagnetic Environ-
ment Guidance
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
GST3C transmitter, including ca-
bles, than the recommended
separation distance calculated
from the equation applicable to
the power of the transmitter.
Refer to the recommended sep-
aration distance table for more
information.
Conducted RF
IEC 61000-4-6
3 V/m
150 kHz to
80 MHz
Not applicable Not applicable