User's Manual
To remove adhesive residue, you will need the following materials: Detachol
®
medical adhesive remover and cotton swabs.
To remove adhesive residue:
1
Make sure the tester is attached to the transmitter.
2
Holding the tester, saturate a cotton swab in the Detachol solution and gently rub
the adhesive residue on the transmitter until it is fully removed.
3 Continue with the cleaning procedure above.
Cleaning the charger
This procedure is for general cleaning as required, based on physical appearance.
Caution: The charger is NOT waterproof. Do NOT immerse in water or any
other cleaning agent.
Caution: Dispose the charger according to the local regulations for battery
disposal (non-incineration).
To clean the charger:
1
Wash your hands thoroughly.
2
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent,
to clean any dirt or foreign material from the outside of the charger. Never use
organic solvents, such as paint thinner or acetone, to clean the charger.
3
Place the charger on a clean, dry cloth and air dry for 2–3 minutes.
Troubleshooting
The following table contains troubleshooting information for the transmitter, charger,
and tester. For more information about troubleshooting, see your pump user guide.
-13-
English
MP6025977-021 / 1