User's Manual
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Guardian Link is intended for use in the electromagnetic environment specified
below. The customer or the user of the Guardian Link should assure that it is used in
such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Envi-
ronment - Guidance
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 400 A/m
4000 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical lo-
cation in a typical domestic,
commercial, or hospital
environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Guardian Link is intended for use in the electromagnetic environment specified
below. The customer or user of the Guardian Link should assure that it is used in such
an electromagnetic environment.
Immunity Test IEC 60601 Level Compliance Lev-
el
Electromagnetic Environ-
ment Guidance
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
Guardian Link, including cables,
than the recommended separa-
tion distance calculated from the
equation applicable to the power
of the transmitter.
Refer to the recommended sep-
aration distance table for more
information.
Conducted RF
IEC 61000-4-6
3 V/m
150 kHz to
80 MHz
Not applicable Not applicable
-16-