3 Hold the MiniLink as shown at right.
Touch the inserted sensor at back of
assembly to prevent movement..
4 Align the MiniLink connector with the
sensor connector. The MiniLink flat
underside (with the label) should face
the skin.
5 Slide the MiniLink onto the sensor.
Make sure it is pushed in firmly.The
flexible side arms of the sensor will “click”
into notches on the underside of the
MiniLink. The MiniLink will flash green for
about 10 seconds with a proper
connection.
6 The MiniLink will begin to send signals to
the monitor or the pump (see your pump
or monitor user guide for instructions).
Note: If you connect the MiniLink to a sensor that has not been inserted into