If the charger AAA battery is low AND the MiniLink battery is bad, the red LED will
blink rapidly for 1 second, pause for 1 second, then blink once. After a 2 second
pause, the cycle repeats. Call the Medtronic Diabetes 24 Hour HelpLine.
Setting up the MiniLink transmitter
1 Find the MiniLink 7-digit serial number on the flat
side of the device as shown at right. The serial
number starts with “2”.
2 Enter the serial number into your continuous
glucose sensing system monitor or sensor-
enabled pump. See your monitor or pump user
guide for using a sensor transmitter with your
system.
Connecting to sensor
1 Follow the instructions provided with your sensor
and insert the sensor before connecting the MiniLink.
2 Once the sensor is inserted and adhered to your body, wait 5 minutes to check
for bleeding. Make sure that any bleeding has stopped (see Warnings page for
details), then connect the MiniLink to the sensor.