Alarm Snooze - Once a sensor alarm occurs the
pump will not repeat the alarm until after this
period of time. This is the setting for the Meter
BG Now alarm.
B
BG Units - Blood glucose units used by the pump
(mg/dL or mmol/L). The BG units can only be
set from the sensor “Edit Settings” screen if the
Bolus Wizard is turned off.
C
Cal Reminder - The pump will trigger a Meter BG
Now alarm automatically every 12 hours,
signaling that the current calibration value is
no longer valid. The value of the Cal Reminder
is the amount of time before the current
calibration value expires when the user wants
to be reminded to calibrate by having the pump
issue a Meter BG Now alarm. For example, if
the Cal Reminder is set to 2 hours, the Meter
BG Now alarm will occur 2 hours before the
calibration is required.
H
High Glucose - The pump will alarm if the sensor
indicates that the user’s blood glucose is at or
above this value. You have the option to turn
this feature on or off.
High Snooze - Allows the user to set the delay
between the first High Glucose Alarm and any
subsequent alarms. This will allow the user to
avoid an alarm every five minutes until the
condition is corrected.
L
Low Glucose - The pump will alarm if the sensor
indicates that the user’s blood glucose is at or
below this value. You have the option to turn
this feature on or off.
Low Snooze - Allows the user to set the delay
between the first Low Glucose Alarm and any
subsequent Low Glucose Alarms. This will allow
the user to avoid an alarm every five minutes
until the condition is corrected.
M
Missed Data - The pump will alarm if it has not
received data from the sensor for an amount of
time that you set.
P
Pump S/N - Pump S/N is the serial number of the
pump currently in use.