User Manual

Troubleshooting and alarms 35
Meter BG now
A meter BG is needed right away to update sensor and to keep receiving
sensor glucose data.
Meter BG by
A meter BG entry is required by the time that is shown to update sensor
and to keep receiving sensor glucose data.
Low mg/dl
The glucose value is lower than or equal to the low glucose limit set. If
you do not set a Low BG then you will not get a Low BG alarm.
High mg/dl
The glucose value is higher than or equal to the high glucose limit set. If
you do not set a High BG then you will not get a High BG alarm.
8:35A
METER
BG NOW
Sensor reading
invalid Enter
meter BG now
ESC, ACT to clear
METER
BG BY 7:23 p
This is a reminder to
enter meter BG soon
ESC, ACT to clear
_:_
12:05A
LOW
40 MG/DL
Glucose is lower than
user specified limit
ESC, ACT to clear
11:17A
HIGH
200 MG/DL
Glucose is higher than
user specified limit
ESC, ACT to clear