User Manual

Troubleshooting and alarms 33
Sensor alarm conditions
Listed below are the alarms that you may encounter while using the sensor feature of your pump,
along with how to resolve the alarm condition.
Weak signal
Alerts you when the pump does not receive valid data for a period of
time, as set in Missed Data. Move the pump to a new location on your body.
Lost sensor
The pump has not received a signal from the transmitter for more than 40
minutes. Make sure the transmitter and sensor are connected. If you hear a
beep you will have to recalibrate. Do NOT disconnect. To find your sensor
use the Find Lost Sensor function:
Main Menu > Sensor> Sensor Start > Find Lost Sensor
Low transmtr
Occurs when the transmitter battery is low. Replace transmitter. Need to
order new transmitter.
9:42A
Select Patterns
WEAK
SIGNAL
Sensor too far
away from pump
See user guide
ESC, ACT to clear
11:17A
LOST
SENSOR
Pump no longer
getting Sensor
data. See user
guide ESC, ACT to
clear
12:00A
LOW
TRANSMTR
Replace
transmitter now
ESC, ACT to clear