User Manual

Programming your sensor 9
High snooze feature
Once you get a High Glucose Alarm, the pump alarms about every 5 minutes while the high glucose
condition exists, unless you set the High Snooze feature. This allows your glucose to return to normal
after you take some insulin without having the pump alarm about every 5 minutes. You can set the
time period for between 5 minutes to 3 hours.
Low glucose alarm
Your pump will alarm if your BG reaches or goes below what you set here. If you do not turn on the
Low Glucose Alarm, your pump will not alarm when your BG goes low.
6 Press ACT to program
the High Snooze feature.
7 Use the up and down arrow
buttons to select your High
Snooze alarm time and press
ACT.
8 Press ACT to program your
Low Glucose limit.
9 Select ON and press ACT. 10 In the SET LOW GLUCOSE screen, use the up and
down arrow buttons to set your low glucose limit and
press ACT.
-
EDIT SETTINGS
Sensor: On
High Glucose: 200
High Snooze: 1:00
ACT
(flashing) set, then
press ACT
SET HIGH SNOOZE
1:00
ACT
EDIT SETTINGS
High Snooze 1:00
Low Glucose: Off
Alarm Snooze: 0:05
ACT
LOW GLUCOSE
Off
On
ACT
(flashing) set, then
press ACT.
SET LOW GLUCOSE
50mg/dL
ACT