User Manual
Troubleshooting and Alarms 73
Off/No Power
Error Message: OFF/NO POWER (Replace Batteries)
Cause: There is less than one (1) hour of battery life left. The Monitor will turn off.
Action:
1. Press the SEL button and then the ACT button to CLEAR the alarm.
2. CHANGE THE BATTERIES RIGHT AWAY TO AVOID LOSS OF DATA STORED IN MEMORY! Follow steps
in Chapter 2: Getting Started.
3. Perform the manual Search procedure:
a. From the HOME Screen, press the SEL button 5 times until the SEARCH screen appears.
b. Press the ACT button and the Monitor will beep or vibrate. CONFIRM will be blinking on the
screen.
c. Press the ACT button again to begin a SEARCH. The screen will display SEARCH and the number 8
(minutes). A countdown will begin. It may take up to 8 minutes to finish the SEARCH.
12:00
AM
NO POWER
-
+
OFF