User Manual

Getting Started 47
Connecting the Sensor to the Transmitter
NOTE -
To extend Transmitter battery life, do not leave Sensor or Test Plug connected to the
Transmitter when system is not in use.
1. With the Monitor in SEARCH mode, hold the base of the Sensor
between the thumb and forefinger so that it doesn’t move. Hold
the Transmitter connector in the other hand. Place the Sensor
connector so that the two connectors are closely aligned. Push
together to connect the Transmitter to the Sensor. You should
hear a click when fully connected.(See Figure 2.5.)
Figure 2.11
2.
The Monitor will search for signals sent by the Transmitter. If the Transmitter is not identified by
the Monitor, an alarm code will appear. See Chapter 4: Troubleshooting and Alarms for detailed
information on Alarm codes.
3. When the Transmitter is identified by the Monitor, the screen will briefly display SUCCESS. The INIT
screen will then appear.
4. Cover the Sensor site with a sterile/snug dressing, such as
IV3000
®
. (See Figure 2.6.)
Figure 2.12
NOTE -
If the Transmitter is unplugged from the Sensor, wait 20 seconds before reconnecting in
order to ensure normal operation
SUCCESS
3:58
PM