User Manual
44 Chapter 2
Sensor Storage and Handling
➠ Store Sensors in refrigerator at +36° to +50°F (+2° to 10°C). Do not freeze.
➠ Before opening, allow individual Sensor package to reach room temperature and 5% to 95% relative
humidity to avoid condensation.
➠ Sensors taken out of refrigeration must be kept at or below room temperature, +50° to +75°F (+10°
to 24°C), and used within 1 week to ensure Sensor sensitivity.
➠ Throw away Sensor after “Use Before” expiration date on label, or if the package is damaged or the
seal broken.
➠ Do not use Sensor if High temperature-limit indicator is black (not clear).
Sen-serter Maintenance
➠ Occasionally clean Sen-serter with soapy water, using liquid detergent or other household soaps.
Allow to dry.
➠ Disinfect Sen-serter by wiping with 10% bleach solution or 70% isopropyl alcohol as needed.
➠ Always store Sen-serter in the released position to ensure peak product performance and life.