Sensor Storage and Handling
Store Sensors in refrigerator at +36° to +50°F (+2° to 10°C). Do not freeze.
Before opening, allow individual Sensor package to reach room temperature and 5% to 95% relative
humidity to avoid condensation.
Sensors taken out of refrigeration must be kept at or below room temperature, +50° to +75°F (+10°
to 24°C), and used within 1 week to ensure Sensor sensitivity.
Throw away Sensor after “Use Before” expiration date on label, or if the package is damaged or the
seal broken.
Do not use Sensor if High temperature-limit indicator is black (not clear).
Sen-serter Maintenance
Occasionally clean Sen-serter with soapy water, using liquid detergent or other household soaps.
Allow to dry.
Disinfect Sen-serter by wiping with 10% bleach solution or 70% isopropyl alcohol as needed.
Always store Sen-serter in the released position to ensure peak product performance and life.