User Manual
Getting Started 43
9. While holding Sensor in place, gently slide Sen-serter away from Sensor. Do not twist, bend or lift
Sen-serter while taking it off Sensor.
10. Holding Sensor base, take off white paper from adhesive pad. Press adhesive against skin.
11. Hold Sensor with two fingers on base, and gently take out introducer needle at 45-degree angle.
➠ Do NOT rotate introducer needle when taking it out
➠ Throw away needle in sharps container
➠ Check site for redness, bleeding, irritation, pain, tenderness or swelling
NOTE -
The expected life of the Sensor is at least 24 hours and is good for up to 72 hours of use.
Replace Sensor at 72 hours, or earlier in the event of a REPLACE SENSOR (C80) alarm. If these
alarms are triggered, follow instructions in Chapter 4: Troubleshooting and Alarms.