Setting the Low Glucose Alert
The Monitor is designed to alert the user when the system finds glucose values at or above the set high
glucose limit, or at or below the set low glucose limit. These alerts may be turned off if not needed or
wanted. If an alert is triggered, you should confirm the glucose value with a home glucose meter before
taking any action. Also, you should use a home blood glucose meter to confirm the symptoms of high and
low blood glucose. (See Chapter 1: Introduction for a list of symptoms.)
The low glucose alert can be set from 40-100 mg/dl (2.2-5.5 mmol/l). Enter a glucose value within this
range to set a low glucose alert. Typically, the number entered is above the value that causes you to have
symptoms of low blood glucose. Confirm the alert limit with your doctor.
For this feature to work, you must select ON, a value, and CONFIRM to save the value in memory. Always
treat low blood glucose per the advice of your doctor and the readings of your blood glucose meter.
From the HOME screen:
1. Press the SEL button 7 times until the SETUP screen appears.
2. Press the ACT button once to go to the SETUP sub-screens.
3. The LOW Limit sub-screen will automatically appear with the previously set low-glucose limit (or the
default value of 60 mg/dl (3.3 mmol/l) ) showing in the upper right hand of the display.
4. Press the ACT button and the following screen appears with ON (default setting) blinking:
LOW
LIMIT
GLUCOSE
60
mg/dl
LOW
LIMIT
GLUCOSE
ON
SET