User Manual

116
Table A2
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Guardian RT is intended for use in the electromagnetic environment specified below. The
user of the Guardian RT should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+ 6 kV
+
8 kV air
> +
8 kV
> +
30 kV air
Typical commercial,
hospital, or home
environment. Synthetic
materials may be
present. The relative
humidity may be as low
as 10%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
Not Applicable Not Applicable Not Applicable
Surge
IEC 61000-4-5
Not Applicable Not Applicable Not Applicable
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
Not Applicable Not Applicable Not Applicable
Power Frequency
(50-60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.