User Manual

104 Chapter 6
Cleaning the Sen-serter
For single-patient use
1. Dampen a clean cloth with mild liquid soap and warm water and wipe the sen-serter.
2. Rinse with warm tap water.
3. Using an antibacterial hand-sanitizer (readily available at your local drugstore), wipe down the
sen-serter.
4. Place the sen-serter on a clean dry cloth and allow to air dry.
5. Store sen-serter in the released position to maintain optimum product performance and life.
For multiple-patient use
1. Dampen a clean cloth with mild liquid soap and warm water, and wipe the Sen-serter.
2. Rinse with warm tap water.
3. Apply 3-4 drops of quaternary ammonium compound disinfectant (e.g., CaviCide
®
) on a clean, dry
cloth. Wipe down the sen-serter.
4. Using 70 percent isopropyl alcohol, or isopropyl alcohol wipes, wipe down the sen-serter.
5. Place the sen-serter on a clean dry cloth and allow to air dry.
6. Once the sen-serter is dry, place it in a sealed bag, labeled with the cleaning date. (See example
below.)
Example: Multiple-use cleaning label
Device
:
Date:
Method of Decontamination
Wash:
Disinfectant: