User's Manual

5
communication problems can typically be resolved by ensuring that the distance between
transmitter and receiving device is less than 1.8 meters (6 feet), and by turning o or moving
away from other RF transmitting devices. You can also reorient or relocate the CareLink USB and/
or the transmitting device to try to correct the interference. Note that RF interference will not
cause any incorrect data to be processed and will not cause any harm to your CareLink USB.
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject
to the following two conditions: (1) this device may not cause interference, and (2) this device
must accept any interference, including interference that may cause undesired operation of the
device.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio
exempts de licence. Lexploitation est autorisée aux deux conditions suivantes : (1) l’appareil
ne doit pas produire de brouillage, et (2) l’utilisateur de l’appareil doit accepter tout brouillage
radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement.
Wireless coexistence summary
The CareLink USB (MMT-7306) was tested against the data listed in the two tables for wireless
phones and wireless networks. The test results are described after the two tables.
Wireless phones
Telephones Frequency
Recommended Separation Distance
(meters)
Recommended Separation Distance
(inches)
WCDMA
800 - 960 MHZ,
1700-2200 MHZ
0.3 12
TDMA 0.3 12
GSM 0.3 12
DECT 0.254 10