User's Manual

3
Connecting the CareLink USB to your computer
Plug the CareLink USB into your computer when prompted by the CareLink
software. It is recommended that the CareLink USB be plugged directly into a
USB port on your computer, rather than into a peripheral device, such as a USB
port on your monitor or keyboard. You can use a USB extension cable if you do
not have easy access to the USB ports on your computer, but it is recommended
that you use only a USB cable with a protected female end (see image).
Troubleshooting
If you plug your CareLink USB into a USB port on your computer and it does not work,
remove it and plug it into another USB port on your computer. If your CareLink USB does
not work in any USB port on your computer, call the HelpLine for assistance or your local
Medtronic representative.
Electrostatic discharge (ESD) may temporarily prevent the CareLink USB from functioning. If
you touch the metal components of the CareLink USB or the uncovered end of a USB cable
while it is connected to your computer, and your CareLink USB stops transmitting, remove it
and plug it back in, being careful not to touch the metal parts.
If you are a healthcare professional, see the CareLink Pro User Guide for more information
about supported devices, troubleshooting, or other instructions.
This device cannot be used to upload pumps other than the MiniMed 600-series pumps.
Compliance information
This device complies with the United States Federal Communications Commission (FCC) and
international standards for electromagnetic compatibility. The CareLink USB (MMT-7306)
conforms to the essential requirements of the Council Directive 1999/5/EC (R&TTE).
The following information is provided in accordance with Federal Communications Commission
(FCC) regulations.