User's Manual

8
Table 2. Guidance and Manufacturers Declaration - Electromagnetic Immunity
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specied below. The customer or the
user of the CareLink USB should make sure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Electrical fast
transient/burst
±2 kV for power
supply lines
Not applicable
For use in a typical domestic, commercial, or hospital environment.
IEC 61000-4-4
±1 kV for input/
output lines
±1 kV for
input/output
lines
Surge ±1 kV line(s) to
line(s)
±1 kV line(s) to
line(s)
IEC 61000-4-5 ±2 kV line(s) to
earth
±2 kV line(s) to
earth
Voltage
dips, short
interruptions
and voltage
variations on
power supply
lines
IEC 61000-4-11
0% 0.5 cycles
0% 1 cycle
70% 25 cycles
0% 250 cycles
0% 0.5 cycles
0% 1 cycle
70% 25 cycles
0% 250 cycles
Power
frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Table 2, continued next page