User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
Using Bolus Wizard 65
5. Review the information in the ESTIMATE DETAILS screen.
Press ACT to continue to step 6. If you need to make any
changes, press ESC to return to the ENTER BG screen (step 3)
and make changes as necessary.
6. In the SET BOLUS screen, the estimated bolus amount will
show (flashing). Change the amount if desired. Press ACT
to accept.
7. Press ACT to accept and deliver the bolus. The BOLUS
DELIVERY screen will appear. The pump will beep or vibrate at
the start and end of the bolus. As the bolus is delivered, the
screen will show the bolus type and amount until the total units
have been delivered. The screen will then default to the HOME
screen.
4. In the ENTER FOOD screen, if this is a food bolus: if this is a correction bolus:
enter the food value you
will eat and press ACT.
select 0 (zero) as the value and
press ACT.
MiniMed
ENTER FOOD
grams*
45
(flashing)
* or exch
MiniMed
ENTER FOOD
grams*
0
(flashing)
* or exch
Food intake: 45gr
(Meter) BG: 160
Food: 3.0U
Correction: 2.0U
Active Ins: 1.0U
Est total: 4.0U
(values shown are
for example only)
ESTIMATE DETAILS
ACT to proceed,
ESC to back up
ACT
to continue
ESC
returns to the
ENTER BG screen
MiniMed
SET BOLUS
4.0U
ACT
to accept
MiniMed
BOLUS DELIVERY
0.
1U
3:24P
NORMAL