User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
64 Chapter 5
Normal bolus using Bolus Wizard
After the Bolus Wizard is turned on and the settings are programmed, Bolus Wizard can calculate an estimate of
insulin you need for your correction bolus and/or your food bolus. You have the option of using the estimate or
changing it as necessary. Additionally, your pump can receive your BG reading from the Paradigm Link meter, if
they are linked.
Use to deliver a Normal bolus at any time except during another Normal bolus. A Normal bolus will temporarily
interrupt a Square Wave or Dual Wave bolus that is delivering. After the Normal bolus is finished, the Square Wave
or Dual Wave bolus delivery will resume.
NOTE - If you want to use the pump-to-meter link, make sure the meter option is on. Refer to the
section, “Meter option” for instructions.
1. If you want a correction bolus, check your BG with your BG meter and go to step 2.
If you want to bolus for food, go to step 2.
2. Press on your pump, or go to the BOLUS MENU, select Use Bolus Wizard, and press ACT.
3. The ENTER BG screen will appear.
If you are NOT using the Paradigm Link meter: If you are using the Paradigm Link meter:
Enter your BG value.
Press ACT and continue to step 4.
If you are not entering a BG
and want to bolus for food,
select the dashes in the
ENTER BG screen. Press ACT
and continue to step 4.
NOTE - Selecting dashes in this screen will
make Bolus Wizard calculate the insulin
needed for your food entry without
considering your BG level.
The meter reading will flash
on the pump screen. Press
ACT to accept this amount.
(You can change this BG
value, if necessary.) Continue
to step 4.
NOTE - You must program your bolus within 12
minutes of the pump receiving the reading from
the meter. If more than 12 minutes have
passed, the reading will no longer be available
from the screen and you must enter your BG
manually.
MiniMed
- - - mg/dL*
ENTER BG
*or
mmol/L
MiniMed
ENTER BG
XXX mg/dL*
meter
*or
mmol/L