User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
54 Chapter 5
How the Bolus Wizard works
More about Bolus Wizard...
About high or low BG levels...
If your BG is less than 70 mg/dL (3.9 mmol/L), the screen
will notify you and give instructions. Bolus Wizard
programming will not continue. Read the instructions and
press
ACT
or
ESC
to clear the message.
The BOLUS
MENU will appear. Exit the menus.
If your BG is more than
250 mg/dL (13.9 mmol/L)
, the
screen will notify you and give instructions.
Read the
instructions and press
ACT
or
ESC
to clear the message.
You can continue programming and deliver your bolus with
Bolus Wizard.
If you are going to eat and you want
Bolus Wizard to estimate your food bolus: Enter your food intake (carbs or exch)
and/or
If you believe your BG is high and you want
Bolus Wizard to estimate your correction bolus: Enter your BG Reading
-
automatically from the Paradigm Link meter
(refer to the section, “Meter option”)
or
-
manually
NOTE - If you are using the Paradigm Link meter, you
can program your pump to automatically receive your
meter readings. The Bolus Wizard will use the BG
reading when calculating your bolus amount. Refer to
the section, “Meter option” in this chapter for
instructions. (If you are not using this meter, you will
manually enter your BG.)
MiniMed
LOW BG
DO NOT BOLUS UNTIL
BG IS NORMAL.
Treat low BG.
Monitor BG.
ACT or
ESC
MiniMed
BOLUS MENU
MiniMed
HIGH BG
Check for occlusion.
Check ketones.
Consider insulin
injection.
Monitor BG.
ACT or ESC
Use Bolus Wizard
Manual Bolus