User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
46 Chapter 4
Fixed prime
A fixed prime fills the soft cannula with insulin and is required after the infusion set is inserted into your body. You
will start the fixed prime after the introducer needle is removed. Additionally, a fixed prime is required if you
disconnect your quick release and you need to reprime the infusion set or if you change your infusion set without
changing the reservoir.
NOTE - Prime amounts depend on the type of infusion set you are using. Refer to your infusion set
instructions for your fixed prime amount.
1. Go to the PRIME MENU.
Main > Prime
2.
Select Fixed Prime and
press ACT.
3. In the FIXED PRIME screen,
enter the amount for your type
of infusion set, then press ACT.
4. Once the prime begins, the PRIME
DELIVERY screen will count up the
units as they are delivered. A “beep”
will sound when priming is complete.
PRIME MENU
Fixed Prime
Prime History
MiniMed
Rewind
ACT
MiniMed
FIXED PRIME
0.
U
3
ACT
MiniMed
6:26P
U
PRIME DELIVERY
0.
2