User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
42 Chapter 4
5. Press ESC. Your manual prime is complete.
6. You can now insert the infusion set into your body as described in the next section.
Insert the infusion set
After you complete all of the following, you will be ready to insert the infusion set into your body:
➠ fill your reservoir
➠ rewind your pump
➠ confirm your insulin type and
➠ prime the pump (fill the infusion set with insulin)
It is important that you change your infusion set every 2-3 days. Medtronic Minimed offers a number of different
infusion sets for your pump. Instructions for the Quick-set® are provided. For other infusion set types, refer to
their user guides. After your infusion set is inserted, continue to the section, “Fixed prime.”
Sof-set® infusion set
Refer to the Sof-set user guide for instructions.
Silhouette® infusion set
Refer to the Silhouette infusion set user guide for instructions.
If your manual prime uses more than 30U insulin, the pump will ask you,
“is priming complete?”
If you get this message, do these steps:
a. Make sure you are not connected to the pump.
b. Read the message on the screen, then press ESC, ACT to clear.
c. If your manual prime is complete (you see droplets on the tip of the infusion set
needle), press ESC and continue to step 5. If your manual prime is not complete,
press and hold ACT until priming is complete then continue to step 5.
WARNING: While the infusion set is connected to your body, do not unscrew and retighten the tubing
connector on the reservoir.
IS PRIMING
Disconnect from body.
COMPLETE?
See user guide
for instruction.
ESC, ACT to clear