User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
40 Chapter 4
Inserting the reservoir in your pump
If your reservoir is already inserted in your pump, continue to the next section, “Manual prime.”
You must do these steps in the order described. Your pump screen will show instructions to help you with these
steps. If you are practicing, do NOT insert the reservoir in your pump.
1. If you are using the pump for the first time, remove the shipping cap from the reservoir compartment.
4. Attach the activity guard, if desired.
5. You must now do a manual prime as described in the next section.
CAUTION: You must rewind your pump before installing a new reservoir. As part of the pump’s function,
it measures the reservoir volume. To ensure correct volume measurements, your pump has been
designed to require a rewind after you change your reservoir.
WARNING: Do not insert the reservoir in the pump if you did not rewind. Doing so could result in
inaccurate insulin delivery.
Never insert the reservoir into the pump while the tubing is connected to your body.
Doing so could result in an accidental infusion of insulin.
2. Insert the reservoir into the top of the pump case. 3. Turn the tubing connector approximately
1/2-turn clockwise until the connector is seated.
The tubing connector should be aligned
horizontally with the pump case as shown here.
512 pump shown;
712 pump similar
tubing connector